Macros: PM200 and PM600 - Fully Automated Calibration

See the attached files. Subscribe to receive emails of updates to this article (and files).

 

This article describes the process of automating the calibration of PM200 and PM600 pressure modules using COMPASS for Pressure and the "Cal Sled" (PMM-CAL-KIT-20M).   

The automated process requires the use of COMPASS macros.  If your database (typically named "compsetp.mdb") does not have the macros you will need to add them.  MS Word files are provided to copy the VB Script code and paste into the COMPASS via Macro Editor tool.

The automated process consists of taking As Received data, Making an adjustment to the calibration coefficients, and then finishing with an As Left verification run.  

A pair of example databases are provided -- one for users who are running COMPASS for Pressure version 4.2 or older, and one for those running version 5 or newer.

Have more questions? Submit a request

Comments

  • Avatar
    Matt Daniels

    If you use the example databases, it is recommended that you import them into your regular database by using the COMPASS Database Maintenance Tool. Then you will need to update the specifics of the Test and DUT definitions such as range, serial number(s), COM ports, and the READ/SET standards used for the testing.