Precision, Performance, Confidence. ™

Error 4 on a Drywell

Tech Notes

          

--------------------------------------------------------------------------------------------------------------------------------

Title: Error 4 on a Drywell

Created: 13 Apr 2012

Last Revised: 13 Apr 2012

--------------------------------------------------------------------------------------------------------------------------------

 

 

Applies To: All Drywells

 

Problem Description: Error 4 is displayed on the drywell

 

Resolution/Work Around:  The error 4 is an ADC error (Analog to Digital Converter).  This is caused when the ADC is not converting or producing a usable signal to the controller.  The ADC chip has gone bad.  Replacement of the analog board or the controller, dependant on the model being serviced, is recommended.