Pressure Measurement Module (PMM) Calibration Kits (and Commands to use with)

Pressure Measurement Module (PMM) Calibration Kits (Calibration Sled) are available if you would like to calibrate PM600, PM630, PM500, PM200 or PM230 modules outside of the pressure controller. The kits include a power supply adapter, a null modem cable (DB9-F to DB9-F), and a null modem USB to RS232 adapter cable (and driver CD).

The TEST and REFERENCE pressure ports of the 20M and 40M kits are 1/4 inch NPT-F, and are DH500-F on the 100M kit. DH500 is a cone and threaded connection compatible with Autoclave F250C and HIP  HF4.

COM settings are: baud rate 9600, parity none, data bits 8, stop bits 1, handshake none. Commands sent to the Cal Sled should be terminated with CR and LF. The replies from the Cal Sled are terminated with CR. Here is a view of these settings in the Fluke Calibration Remote Communications Tool that is included with COMPASS for Pressure calibration software. The settings in HyperTerminal, PuTTy, or NI Interactive Control.

Note that communications with the modules when they are installed in a pressure controller (2271A, 6270A, 8270A, 8370A) are by CAN bus so the below remote commands should not be sent to pressure controllers, just to the modules when they are in a calibration kit.

Part Number: 4968763
Description: PMM-CAL-KIT-40M, PRESSURE MEASUREMENT MODULE CALIBRATION KIT, 40 MPA (6000 PSI).
Notes: For use with all modules 40 MPa (6000 psi and below). This kit comes with a plug for the REFERENCE port for use when calibrating a BRM600-BA100K module. It is recommended to purchase this kit rather than the -20M model as it will calibrate all modules that the -20M kit will, and it calibrates modules up to 40 MPa also.

Part Number: 4456687
Description: KIT-PMM-CAL-20M, PRESSURE MEASUREMENT MODULE CALIBRATION KIT, 20 MPA (3000 PSI).
Notes: For use with all modules 20 MPa (3000 psi and below). This kit comes with a plug for the REFERENCE port for use when calibrating a BRM600-BA100K module.

Part Number: 4968756
Description: KIT-PMM-CAL-100M, HIGH-PRESSURE MEASUREMENT MODULE CALIBRATION KIT, 100 MPA (15000 PSI).
Notes: For use only with PM230 and PM630 modules - modules that are 70 MPa (10000 psi) and higher. 

If you need the driver for the USB-RS232 adapter see this article, USB to null modem RS232 adapter

To order or for a quote, contact the Fluke Calibration Orders Team by email at FCAL Orders Team or by calling +1.877.355.3225. Contact the Fluke Calibration Pressure Technical Support team at the link below (to create a support ticket) or at the same phone number. 

Note: Commands below are in quotes for clarity of what the full command is. Don't send quotation marks. 

PRES_UNIT?
To read the pressure unit for the VAL? query send "PRES_UNIT?" command. The return is the pressure unit.
To set the pressure unit for the VAL? query send "PRES_UNIT x" where "x" is KPAL (for kPa), PSI or BAR

VAL? or VAL1?
To read the current pressure send "VAL?" command. The reply will be the value in the unit specified by PRES_UNIT command.

VAL2?
To read the module's on-board barometer current pressure (if applicable) send "VAL2?" command. The reply will be the value in the unit specified by PRES_UNIT command. This command is not needed for calibrating the module's pressure output.

VAL3?
To read the module's fast sensor current pressure (if applicable) send "VAL3?" command. The reply will be the value in the unit specified by PRES_UNIT command. This command is not needed for calibrating the module's pressure output.

TEMP?
To read the current temperature send "TEMP?" command (if a temperature output is available). The reply will be the value and current units. This command is not needed for calibrating the module's pressure output.

*IDN?
To read the serial number send "*IDN?" command. Reply is "Fluke, Model #, Serial #, software version, model, memory. Example reply is FLUKE,PM600-A20M,9990101,1.00,PM600,64K

Whenever you make changes to settings you must send “CAL_STORE” to save to memory. If not, and power is cycled, the value will not be stored.

C0 (Pressure Adder)
To read C0 send “OFFSET_SET?” The return is always in psi.
To set C0 send “OFFSET_SET x” where “x” is a decimal or integer and must be in psi.

Note - On the pressure controller display screens, the CO value is in psi for PM200, PM600 & PM630 modules. The C0 value is in kPa for PM500 modules (as of 12-December-2018, will likely be changed to psi sometime in the future).

C1 (Pressure Multiplier)
To read C1 send “GAIN_SET?” The return is a unit less decimal.
To set C1 send “GAIN_SET x” where “x” is a decimal (multiplier).

ZOFFSET (zero offset or zOffset or Autozero value)
To read zOffset send “UCOEF_SET[0]?” The return is in psi.
To set zOffset send “UCOEF_SET[0] x” where “x” is a decimal or integer in psi.

Autozero (Zeroing function)
The following user command zeros the test sensor to the reference value given, or can reset the existing zero offset prior to calibrating the sensor. The reference value is in psi.
To read the previously sent reference value send "TEST_ZERO?" This is not the Autozero (zOffset) value but the reference value, and is in psi.
To zero the sensor send "TEST_ZERO ref" where ref is the reference pressure in psi. Set ref to zero (0) for gauge mode sensors.
To clear the zOffset (Autozero) value send "TEST_ZERO RESET"

CAL DATE for the primary sensor
To read cal date send “CAL_DATE?” The return is a string.
To set cal date send “CAL_DATE mm/dd/yy”. Ensure that the format matches that of the reply above.

CAL DATE for the primary sensor
To read cal date send “CAL_DATE?” The return is a string.
To set cal date send “CAL_DATE mm/dd/yy”. Ensure that the format matches that of the reply above.

CAL_STORE
To save changes to memory send "CAL_STORE" command

Whenever you make changes you must send “CAL_STORE” to save to memory. If not, and power is cycled, the value will not be stored.

The PMM Calibration Kit is just like a bay in the pressure controller. Remove the Pressure Measurement Module from the bay in the pressure controller and insert it in the same way into the PMM Calibration Kit.

For advanced use with COMPASS for Pressure see this article,
Macros: PM200 and PM600 - Fully Automated Calibration

Have more questions? Submit a request

Comments

  • Avatar
    Alec Zhou

    What's the difference between 'C0' and 'ZOFFSET'?

    Comment actions Permalink
  • Avatar
    Kyle Clark

    CO is the adder (also called offset or PA) calibration adjustment value. C1 is the multiplier (also called slope or PM) calibration adjustment value. zOffset (called Autozero in the 6270A manual) is the same as C0 but is intended to be adjusted between regular calibrations. After a regular calibration the zOffset value is always zero because the adjustment is done with the C0 value. If configured, the 6270A will automatically adjust the zOffset value of modules.

    Comment actions Permalink
  • Avatar
    Leniewerson Franca

    Vocês já pensaram em desenvolver um software semelhante ao software de comunicação utilizado com o monitor 2456-LEM Monitor para ambiente de laboratório da Fluke.(https://br.flukecal.com/products/humidity-calibration/humidity-data-loggers-and-monitors/2456-lem-monitor-para-ambiente-de-l).

    Edited by Leniewerson Franca
    Comment actions Permalink
  • Avatar
    Leniewerson Franca

    Quando enviado o comando de "TEST_ZERO ref" ele faz o zero do sensor?
    E outra coisa como a comunicação do modulo e o calibrador 2271A é feito pelo barramento Can. A leitura do sensor é enviada por disparo continuo ou o controlador deve solicitar a leitura conforme a comunicação RS232?

    Comment actions Permalink
  • Avatar
    Leniewerson Franca

    Na comunicação RS232 a leitura do comando "VAL?" é uma leitura instantânea ou uma media móvel?

    Comment actions Permalink
  • Avatar
    Kyle Clark

    Hello Franca - See this article this Help Center that shows how to use COMPASS for Pressure calibration software similar to LEMcal software for lab environment. Macros: 2456-LEM - Log to a file with COMPASS for Pressure, and read from the file .

    Also see the 1620A DewK with LogWare for lab monitoring,
    https://us.flukecal.com/products/data-acquisition-and-test-equipment/data-acquisition/1620a-digital-thermometer-hygrometer

    ---
    Translated by Microsoft Bing
    "Have you thought about developing software similar to the communication software used with the 2456-LEM Monitor for Fluke's lab environment. (https://br.flukecal.com/products/humidity-calibration/humidity-data-loggers-and-monitors/2456-lem-monitor-para-ambiente-de-l)."

    Comment actions Permalink
  • Avatar
    Kyle Clark

    Hello Franca - Yes, sending the "TEST_ZERO ref" command zeroes the module to match the ref pressure sent with the command. This is most often done in absolute mode but can also be done in gauge mode. This is done with the TEST and REFERENCE ports vented (open to atmosphere).

    CAN Bus is the communications used when the module is installed in the controller. RS232 is the communications when the module is installed in the calibration kit/sled. The command STREAM PRESF or STREAM OFF can be sent by RS232 to stream the filtered pressure values or just send them when queried.
    ---
    Translated by Microsoft Bing
    "When sent the "TEST_ZERO ref" command does it zero the sensor?
    And another thing like module communication and 2271A calibrator is done by Can Bus. Is the sensor reading sent by continuous firing or should the controller request the read according to rs232 communication?."

    Comment actions Permalink
  • Avatar
    Kyle Clark

    Hello Franca - The reading in response to the VAL? query is an instant read from the module.
    ---
    Translated by Microsoft Bing
    "In RS232 communication is the reading of the "VAL?" command an instant read or a mobile media?"

    Comment actions Permalink